Shanghai Haozhuang Biological Safety Cabinet leads the way in energy saving and environmental protection

Shanghai Haozhuang Instrument Co., Ltd. develops the small spray dryer of LNB brand, which is energy-saving and environmentally friendly, leading the country. Shanghai Haozhuang Instrument Co., Ltd. develops the Hao Zhuang (LNB) brand biosafety cabinet on the basis of independent research and development of core technology of scientific instruments by virtue of its profound understanding and mastery of scientific instrument technology. Professional manufacturing is more secure and affordable!

Use and maintenance of biological safety cabinet:
1. Purpose: To standardize the operation and maintenance of the biological safety cabinet to ensure the normal operation of the equipment to ensure the safety of the operators.
2. Scope: BHC-1300A2 biosafety cabinet used by aseptic laboratory inspectors in a sterile laboratory.
3. Responsibilities: The bacteria inspector is responsible for the normal operation and maintenance of the equipment, and the relevant documents are reviewed by the manager of the quality department.
4. Content 4.1 The safety cabinet should be placed in the primary purification room below 100,000 level. Before operation, turn on the UV lamp and irradiate for 30min, and then move all the items required for this operation into the safety cabinet to avoid frequent destruction of the air flow by the arms through the air curtain; Pollution.
4.2 Turn on the fan for 10 minutes, and then conduct the experimental operation after the air in the cabinet is purified and the airflow is stable. Slowly extend your arms into the safety cabinet for at least 2 minutes to make the airflow in the cabinet stable before operating.
4.3 Nothing unrelated to this experiment shall be placed in the biological safety cabinet. Items should be placed as far back as possible, not blocking the airway opening, so as not to interfere with the normal flow of airflow.
4.4 Avoid cross-contamination during operation. In order to prevent possible splashing of droplets, 75% alcohol cotton balls or small pieces of gauze soaked in disinfectant should be prepared to avoid covering the safety cabinet grille with items.
4.5 During the experimental operation, the glass window cannot be fully opened, and the operator's face should be above the working window. Actions should be gentle and soothing when operating in the cabinet to prevent affecting the airflow in the cabinet.
4.6 The safety cabinet should be regularly tested and maintained to ensure its normal operation. Once the safety cabinet is found to be working abnormally, it should immediately stop working, take corresponding measures, and notify the quality department manager.
4.7 After the inspection is completed, close the glass window, keep the fan running for 10 minutes, and turn on the UV lamp for 30 minutes.
4.8 The safety cabinet should be cleaned and disinfected regularly, and the surface of the worktop and cabinet body can be wiped with 75% alcohol or 0.2% Xinjieer solution; after each inspection work, it should be fully disinfected.
4.9 The items used in the cabinet should be taken out after the disinfection tank is sterilized to prevent the standard strains from being carried out and polluting the environment, causing biological hazards.
4.10 How to use:
4.10.1 The safety cabinet is controlled by LED liquid crystal panel. The function keys included in the control panel are: power switch key, sterilization lamp key, windshield key (fast, slow, no wind), lighting key.
4.10.2 After plugging in the 220V AC power supply, press the power switch shown on the control panel to turn it on and use; when sterilization is required, press the sterilization lamp key; the windshield key has three levels (fast, slow, no wind) Cycle, the default is no wind, press once for slow speed, then press for fast, and return to no wind for the third time, so cycle; usually operate lighting, just press the Ming button.
5. Maintenance of equipment 5.1 After each inspection operation, use 75% alcohol (other fungicides depend on the materials used by the user) to thoroughly clean the surface, side walls, back wall, and windows of the work area inside the safety cabinet. Do not use a bactericide containing chlorine, as it may cause damage to the stainless steel structure of the safety cabinet. At the same time, clean the surface of the ultraviolet lamp and the power outlet. When cleaning the internal area of ​​the safety cabinet, the operator can't enter the safety cabinet except any hands.
5.2 If you have not experimented for a long time, you must clean and maintain it every two weeks. Carry out according to 5.1. Regular cleaning of the stainless steel surface will keep it smooth and beautiful.
5.3 Wipe the external surface of the safety cabinet with a damp cloth every month, especially the front and upper parts of the safety cabinet, and clean up the accumulated dust. And check the reasonable use of all maintenance accessories.
5.4 According to the actual situation, check every quarter or half a year for any physical abnormalities or faults in the safety cabinet, and report to the repair if there are any abnormalities; remove and clean the primary effect filter to prevent the accumulation of dust, which will lead to insufficient air intake and reduce the cleaning effect.
5.5 Every year, a qualified certified technician is required to perform performance certification on the safety cabinet, and replace the UV lamp according to the service life effect of the UV lamp; when the normal adjustment or cleaning of the primary effect air filter still fails to achieve the ideal cross-sectional wind speed , The working voltage of the fan should be adjusted (turn the knob) to achieve the ideal uniform wind speed (adjust the wind speed knob from left to right, slowly from low to high speed, the new workbench should not be adjusted to the highest wind speed)
5.6 When the working voltage of the fan is adjusted to the highest point after 18 months of use, but the ideal wind speed is still not reached, it means that the high-efficiency air filter has too much dust accumulation (the filter hole on the filter material has been basically blocked, and it must be updated in time) , The general life span of the HEPA filter is 18 months; when replacing the HEPA filter, attention should be paid to the model specifications (original manufacturer configuration), according to the arrow wind direction device, and pay attention to the peripheral seal of the filter, absolutely no leakage Leakage occurs.
Our professional team provides technical support for you, our brand is worthy of your trust, do every set of equipment, every project, every link, strictly abide by the credibility and commitment, build quality with quality, Shanghai Haozhuang welcomes you!

Mini Elliptical Trainer

Small Elliptical Machine,Mini Exercise Bike,Elliptical Exercise,Body Rider Elliptical

Zhejiang Kingdom Sports Co.,Ltd. , https://www.zjkingdomsports.com